  Testo 330 LL
  testo 330 LL. The only flue gas Analyser with up to 6 years’ sensor lifetime.
  The testo 330 LL is the professional flue gas Analyser which meets the highest requirements and is able to fulfil any measurement task. The sensors have an extended lifetime of up to 6 years. At least one sensor replacement is saved in the course of a normal working life. Testo furthermore gives 4 years’ warranty without a maintenance contract. Many measurement menus allow professional analyses on heating systems. 500.000 measurement values can be stored in the instrument.
  TÜV-tested according to EN 50379, Parts 1-3.    
 
 
Longlife sensors
Up to 6 years’ lifetime.
Measurement data displayed graphically
As a flue gas matrix and a line diagram
Sensor exchange by the user
Simple, by the user.
Stamp of approval
testo 330 LL is TÜV-tested according to EN 50379, Parts 1-3.
Dilution CO
Automatic dilution from 8,000 ppm.
Easy exchange of probes
By probe coupling.
Fast attachment
Magnetic, to the burner/boiler.
Robust design
Ideal for a technican’s daily work.
Lithium rechargeable battery
Operation with lithium rechargeable battery (1500 mAh) – up to 8 hours running time.
Flue gas measurement
The testo 330 directly measures CO and O 2, and additionally ambient and flue gas temperature.
Draught measurement
This allows the negative pressure in the heating system to be quickly determined.
Differential pressure measurement
With the differential pressure measurement, you monitor the gas supply on gas boilers quickly and easily.
Gas leak detection/gas pipe test
This accessory makes additional measuring instruments for these measurements unnecessary.
Ambient CO measurement
With the ambient CO probe, this measurement is possible parallel to the flue gas measurement.
Ambient CO measurement
With the ambient CO probe, this measurement is possible parallel to the flue gas measurement.
  Measuring range Accuracy ±1 digit Resolution Adjustment
time t90

Temperature

-40 to +1200 °C ± 0.5 °C (0.0 to +100.0 °C)
±0.5% of reading (remaining measuring range)
0,1 °C (-40 to +999,9 °C) 1 °C (remaining meas. range)

Draught measurement

-9.99 hPa to +40.00 hPa ±0.02 hPa o. ±5% of. m.v. (-0.50 to +0.60 hPa)
± 0.03 hPa (+0,61 to +500 hPa)
±1.5% of reading (+3.01 to +40.00 hPa)
(the greater value applies)
0.01 hPa

Pressure measurement

0 to +300 hPa ±0.5 hPa (0,0 - 50.0 hPa)
±1% of reading (50.1 to 100.0 hPa)
±1.5% of m.v. (remaining meas. range)
0.1 hPa

O2measurement

0 to 21 vol.% ± 0.2 vol.% 0.1 vol. % < 20 sec

CCO measurement
(without H2 compensation)

0 to 4000 ppm ±20 ppm (0 - 400 ppm)
±5% of reading (401 to 2000 ppm)
±10% of reading (2001 to 4000 ppm)
1 ppm < 60 sec

CO measurement
(H 2-compensated)

0 to 8000 ppm Display range 8000 to 30,000 ppm (automatic dilution) ± 10 ppm or +/-10% of m.v. (0 to 200 ppm)
± 20 ppm or +/-5% of m.v. (201 to 2000 ppm)
±10% of reading (2001 to 8000 ppm)
1 ppm < 60 sec

Only testo 330-2 LL CO measurement (H2compensated) solid fuel measurement

0 to 30,000 ppm ±100 ppm (0 to 1000 ppm)
±10% of reading (1001 to 30,000 ppm)
1 ppm

Efficiency testing (Eta)

0 to 120 % 0.1 %

Exhaust gas loss

0 to 99.9 % 0.1 %

CO 2 -determination
digital calculation from O2

Display area 0 to CO 2 max ± 0.2 vol.% < 40 sec

Option CO low -measure-
ment

0 to 500 ppm ±5 ppm (0 - 100 ppm)
±5% of reading (101 to 2000 ppm)
±10% of reading (2001 to 3000 ppm)
0.1 ppm < 30 sec

Option NO measurement

0 to 3000 ppm ±5 ppm (0 - 100 ppm)
±5% of reading (101 to 2000 ppm)
±10% of reading (2001 to 3000 ppm)
1 ppm < 30 sec

Ambient CO measurement
(with CO probe)

0 to 500 ppm ±5 ppm (0 to 100 ppm)
±5 % of m.v. (>100 ppm)
1 ppm approx. 35 sec

Gas leak measurement for combustible gases (with
gas leak detector probe)

Display area 0 to 10.000 ppm
CH 4 / C 3 H 8
Signal
Optical display (LED)
Audible alarm via buzzer
< 2 sec

Ambient CO2measurement
(with ambient CO 2 probe)

0 to 1 vol. %
0 to 10,000 ppm
±(50 ppm ±2% of reading) (0 to 5000 ppm) approx. 35 sec

Nolow

0 to 300 ppm ±2 ppm (0 - 39,9 ppm)
±5% of reading (40 to 300 ppm)
0.1 ppm < 30 sec

General technical data

Storage temp.

-20 to +50 °C

Oper. temp.

-5 to +45 °C

Power supply

Rech. batt. block 3.7 V / 2,6 Ah Mains unit 6 V / 1.2 A

Memory

500,000 readings

Display

Colour graphic display with 240 x 320 Pixel

Weight

600 g (excluding battery)

Dimensions

270 x 90 x 65 mm

Warranty

Instrument/probe/gas sensors (O2, CO) 48 months
NO-, COlowsensor 24 months
Thermocouple and rech. battery 12 months